Skip to main content

Table 5 Summary of results (SK enriched group)

From: Diastolic timed Vibro-Percussion at 50 Hz delivered across a chest wall sized meat barrier enhances clot dissolution and remotely administered Streptokinase effectiveness in an in-vitro model of acute coronary thrombosis

No.

Donor

Treatment

Lytic Dist

Pre-Wt.

Post Wt.

(%)

R.M.S.

Channels

(min)

Mobilized

Fragmented

1

AH

Control

2.1 cm

0.258

0.254

2

+1

No

-

No

No

2

AH

Vibrate

2.0 cm

0.297

0.184

38

+5

Yes

10

No

No

3

HG

Vibrate

2.0 cm

0.295

0.177

40

+5

Yes

10

No

No

4

HG

Control

2.1 cm

0.288

0.280

3

+1

No

-

No

No

5

AH

Vibrate

1.5 cm

0.371

0.206

44

+6

Yes

10

No

No

6

AH

Vibrate

2.5 cm

0.255

0.126

51

+41

Yes

5

No

No

72

HG

Vibrate

1.5 cm

0.352

0.161

54

+5

Yes

10

No

No

8

HG

Control

2.3 cm

0.262

0.246

6

+03

No

-

No

No

9

AH

Vibrate

2.5 cm

0.236

0.061

74

+6

Yes

5

Yes

No

10

AH

Control

1.5 cm

0.453

0.436

4

+1

No

-

No

No

11

HG

Vibrate

1.8 cm

0.303

0.125

59

+5

Yes

5

No

No

12

AH

Control

2.1 cm

0.286

0.280

2

+1

No

-

No

No

13

HG

Control

1.9 cm

0.299

0.290

3

+1

No

-

No

No

14

HG

Control

1.6 cm

0.341

0.338

1

+1

No

-

No

No

15

AH

Vibrate

2.3 cm

0.255

0.133

48

+5

Yes

5

No

No

16

HG

Control

2.3 cm

0.259

0.251

3

+1

No

-

No

No

  1. Results of; initial lytic to clot separation distance (cm), Percent Clot Dissolution (%), Relative Mixing Score (+0 - +7), clot length fluid channel development (Yes or No), time to first observation of clot length fluid channels (min), clot mobilization (Yes or No) and clot fragmentation (Yes vs. No), in view of a 20 minute exposure of remotely administered SK in conjunction with “diastolic” timed LFVP (Vibrate) vs. no LFVP (Control). Donors, AH = Andrew Hoffmann (male), HG=Harjit Gill (female).
  2. 1Diminished mixing score in this sample was likely attributed to the relatively small initial clot size (with a leading edge ~2.5 cm from SK).
  3. 2This sample run developed a significant leak between the connecting line and blue tip connector of the catheter segment which could not be corrected. Only about 10 minutes of effective LFVP therapy was administered. Mixing score was based on maximum observed prior to the leak.
  4. 3Relatively low mixing score in this sample may have also been attributed to a relatively small clot size (with leading edge 2.3 cm from SK).