Skip to main content

Advertisement

Table 5 Performance of four different tests for rivaroxaban monitoring (threshold: 100 ng/mL)

From: Comparison of different laboratory tests in the evaluation of hemorrhagic risk of patients using rivaroxaban in the critical care setting: diagnostic accuracy study

Test Sensitivity (%, 95% CI) Specificity (%, 95% CI) PPV (%, 95% CI) NPV (%, 95% CI)
PT 100 (85–100) 56 (42–69) 48 (41–55) 100
aPTT 100 (85–100) 67 (53–79) 55 (46–64) 100
dRVVs 100 (85–100) 11 (4–22) 31 (29–33) 100
dRVVc 100 (85–100) 16 (8–29) 32 (30–35) 100
  1. PPV positive predictive value, NPV negative predictive value, CI confidence interval