From: Evaluation of the analytical performance of the PC100 platelet counter
Parameter | Sysmex® XP-300 vs. Sysmex® XN-9000 | |
---|---|---|
Estimate | 95% CI | |
Mean difference (× 103/μL) | −1.15% | − 2.52 to 0.00% |
95% Lower LoA (× 103/μL) | − 18.14% | – |
95% Upper LoA (× 103/μL) | 25.49% | – |
Slope | 1.00 | 0.97 to 1.02 |
Intercept | −3.00 | −6.76 to 3.07 |
Correlation coefficient (r) | 0.99 | – |
Nsamples | 102 | – |
Measuring interval (×103/μL) | 21 to 597 | – |