Skip to main content

Table 2 Results of the method comparison for platelet count in whole blood between Sysmex® XP-300 vs. Sysmex® XN-9000

From: Evaluation of the analytical performance of the PC100 platelet counter

Parameter

Sysmex® XP-300

vs. Sysmex® XN-9000

Estimate

95% CI

Mean difference (× 103/μL)

−1.15%

− 2.52 to 0.00%

95% Lower LoA (× 103/μL)

− 18.14%

95% Upper LoA (× 103/μL)

25.49%

Slope

1.00

0.97 to 1.02

Intercept

−3.00

−6.76 to 3.07

Correlation coefficient (r)

0.99

Nsamples

102

Measuring interval (×103/μL)

21 to 597

  1. 95% CI 95% Confidence Interval, Plts Platelets, LoA Limit of Agreement