Skip to main content

Table 1 Mean CAT parameters and %CV measured using CAT-I and CAT-II method

From: Calibrated automated thrombogram II: removing barriers for thrombin generation measurements

TG parameter CAT-I CAT-II P-value
Mean [SD] CV% Mean [SD] CV%
Lag time (min) 2.65 [0.06] 2.4 2.35 [0.04] 1.9 0.0005
ETP (nM.min) 1284 [49.1] 3.8 1402 [68.32] 4.9 0.0147
Peak height (nM) 246.6 [6.64] 2.7 265.9 [9.59] 3.6 0.018
Time to peak (min) 5.26 [0.13] 2.5 4.98 [0.04] 0.8 0.0195
  1. TG thrombin generation; CAT calibrated automated thrombogram; CV coefficient of variation, ETP endogenous thrombin potential