Skip to main content

Table 2 Mean TG parameters and %CV measured using CAT-II and ST®-Genesia

From: Calibrated automated thrombogram II: removing barriers for thrombin generation measurements

TG parameter CAT-II ST®-Genesia P-value
Mean [SD] CV% Mean [SD] CV%
Lag time (min) 2.35 [0.04] 1.9 2.43 [0.08] 3.1 < 0.0001
ETP (nM.min) 1402 [68.32] 4.9 922.8 [14.49] 1.6 0.0006
Peak height (nM) 265.9 [9.59] 3.6 179.9 [8.77] 4.9 0.0002
Time to peak (min) 4.98 [0.04] 0.8 4.58 [0.17] 3.7 0.0015
  1. TG thrombin generation; CAT calibrated automated thrombogram; CV coefficient of variation, ETP endogenous thrombin potential