Fig. 5 | Thrombosis Journal

Fig. 5

From: Laboratory evaluation of a new integrative assay to phenotype plasma fibrinolytic system

Fig. 5

Influence of the processing temperature (A) and a freeze-thawing cycle (B) on GFC. The red dashed line would indicate the absence of difference between the two settings. The blue dashed line represents the mean (m) of the differences between the two settings and the green ones represent the lower (LB) and the upper (UB) bound of 95% confidence interval for mean of the differences. A Influence of the processing temperature on GFC: GFC assessed at 4 °C was considered as the reference (by analogy with ECLT). Differences between GFC assessed at 20 °C and at 4 °C for each subject were plotted against GFC assessed at 4 °C. B Influence of a freeze-thawing cycle on GFC: GFC assessed with a fresh sample before a freeze-thawing cycle was considered as the reference. Differences between GFC assessed after and before a freeze-thawing cycle for each subject were plotted against GFC before a freeze-thawing sample