Identified Edge of Failure¶ | |||||
---|---|---|---|---|---|
Fluorescence Artifact | Edge of Failure Experiment | Thrombinoscope (TS) Software | In-house ORIGIN (OR) Software | CL Shiny App (SH Software) | TPH Normalization Method |
IFE | FVIII-DP spiked with [AMC]: 0, 47, 70, 102, 200 µM | FVIII = 1 IU/mL: Corrected at [AMC] < 115 µM FVIII = 0 IU/mL: Corrected [AMC] < 75 µM | FVIII = 1 IU/mL: Overestimated results at [AMC] > 170 µM FVIII = 0 IU/mL: Corrected [AMC] > 170 µM | FVIII = 1 IU/mL: Successful at all AMC concentrations FVIII = 0 IU/mL: Successful at all AMC concentrations | FVIII = 1 IU/mL: Successful at all spiked [AMC] FVIII = 0 IU/mL: Successful at all spiked [AMC] |
Substrate Consumption* | Reduced ZGGR-AMC at range: 0-836 µM | n/a*†| n/a* | n/a* | n/a* |
Reduced ratio of ZGGR-AFC supplemented to ZGGR-AMC ZGGR-AFC: 0, 200, 400, 600, 700, 750, 800 µM | n/a*†| n/a* | n/a* | n/a* | |
Procoagulant Samples | ATIII-DP ± heparin with range of TF concentrations successfully induced substrate consumption | n/a†| Not corrected at any TF concentration | Successfully corrected at all TF concentrations | Corrected at all TF concentrations but lag time was modulated |