Skip to main content

Table 2 Different types of technologies for hemostasis point-of-care testing.

From: Point-of-Care Testing of Hemostasis in Cardiac Surgery

DEVICE (MANUFACTURER)

MAIN TESTS

PRINCIPLE

TYPE OF DETECTION

Hemochron series (ITC, USA)

ACT, PT, aPTT RxDX system

Magnet position

Optical detection of change in magnet position in a slowly rotating reagent-filled tube

Hemochron Junior II and Signature (ITC, USA)

ACT+ and LR, PT, aPTT

Fluid oscillation

Optical detection of slowing of oscillatory motion in a test channel

ACT II, Hepcon HMS (Medtronic Hemotec, USA)

HRACT, LRACT, HR-HCT, HMS, HEMOSTATUS PT, aPTT

Plunger motion

Optical detection of change in the fall rate of a plunger assembly that is raised and lowered

RapidpointCoag (Bayer, USA)

PT, PT-ONE, aPTT, HMT

Particle motion

Optical detection of slowing of motion of paramagnetic iron oxide particles in pulsating magnetic field

I-STAT (ABBOTT, USA)

Celite-ACT

Electrochemical conversion

Amperometrical detection of an electroactive product from the conversion of thrombin substrate

CoaguChek Plus and Pro (Roche Diagnostics, USA)

PT, aPTT, ACT

Pattern recognition

Optical detection of change in laser interference pattern from slowing cell motion

ACTALYKE (Array Medical, USA)

ACT, MAX-ACT, PT, aPTT

Electromechanical recognition

Two point electromechanical soft-clot detection

TegR Analyzer (Hemoscope, USA)

Thromboelastogram

Cup oscillation

Mechanical or electrical detection in changes of oscillation of cup and pin together

ROTEGR Analyzer (Pentapharm, BMBH)

Thromboelastogram

Shaft rotation

Optomechanical detection of loss of motion of shaft

SONOCLOT (Sienco, USA)

Sonoclot Signature

Impedance on vibrating probe

Impedance detection of changes in viscoelastic properties of celite-activated blood sample

PFA-100 (Dade Behring, USA)

ADP/CT, EPI/CT

Pressure

Detection of change in pressure across porous aperture when platelet aggregates occlude openings

  1. ACT, Activated Clotting Time; ACT+, Activated Clotting Time plus (for high levels of heparin); ACT LR, ACT low range (heparin); aPTT, activated Partial Thromboplastin Time; ADP/CT, Closure Time by activation with adenosindiphosphate; EPI/CT, Closure Time by activation with epinephrine; HMS, Hemostasis Management System; HMT, Heparin Management Test; HRACT, High Range ACT; HR-HCT High Range Heparinase Clotting Test, ITC, International Technidyne Corp; LRACT, Low Range ACT; MAX-ACT, Maximum Factor XII activation ACT; PFA, platelet function analyzer; PT, Prothrombin Time; RxDX, Heparin and Protamine Response tests