Skip to main content

Table 1 Precision analysis of each parameter of thrombin generation curve given by Thrombogram-Thrombinoscope assessed in fresh PRP in the presence of increasing TF concentrations or with out any TF addition. CV: coefficient of variation.

From: Towards a standardization of thrombin generation assessment: The influence of tissue factor, platelets and phospholipids concentration on the normal values of Thrombogram-Thrombinoscope assay

Recombiplastin final dilution in plasma and approximate TF concentration

Lag-time

Tmax

Cmax

ETP

 

intra-assay CV

inter-assay CV

intra-assay CV

inter-assay CV

intra-assay CV

inter-assay CV

intra-assay CV

inter-assay CV

0 (0 pM)

7%

14%

5%

12%

5%

7%

4%

6%

1/2000 (3 pM)

4%

4%

4%

6%

6%

9%

4%

11%

1/1000 (6 pM)

4%

4%

3%

5%

6%

8%

3%

10%

1/500 (12 pM)

5%

6%

4%

6%

6%

8%

4%

5%

1/200 (30 pM)

6%

9%

5%

9%

3%

5%

4%

8%