Skip to main content
Fig. 2 | Thrombosis Journal

Fig. 2

From: Evaluation of the analytical performance of the PC100 platelet counter

Fig. 2

Method comparison for platelet count in whole blood between Sysmex® XP-300 and Sysmex® XN-9000. The platelet count was determined using the Sysmex® XP-300 or the Sysmex® XN-9000. a In the Bland Altman plot, the difference (%) between the Sysmex® XP-300 and the Sysmex®XN-9000 is plotted against the values obtained by Sysmex® XN-9000. Grey line represents the zero-difference line and the bold blue line represents the median difference. b In the Passing-Bablok regression plot, the grey line represents the y = x line and the bold red line represents the regression line. N = 102