Table 4 Assessment of clinical significance of bias in platelet count of whole blood or PRP between the candidate method PC100 platelet counter and comparative reference methods, Sysmex® XP-300 and Sysmex® XN-9000

From: Evaluation of the analytical performance of the PC100 platelet counter

Sample type

Compared methods

Reference method

(X*(× 103/μL))

Candidate method

(Y*(× 103/μL))

Bias

(Y*-X*)

95% CI

Whole Blood

PC100 platelet counter vs. Sysmex XP-300

50

46.2

−3.85

(−7.7%)

−9.87 to 0.50

(−19.7% to −0.3%)

100

96.4

−3.65

(− 3.6%)

−8.39 to − 0.28

(− 8.4% to − 0.3%)

600

598.4

−1.62

(0.3%)

− 12.99 to 8.04

(−2.2% to −4.7%)

PC100 platelet counter vs. Sysmex XN-9000

50

43.9

−6.13

(−12.3%)

−13.21 to −2.35

(26.4% to − 4.7%)

100

93.9

−6.05

(− 6.1%)

− 11.82 to − 2.59

(− 11.8% to − 2.6%)

600

594.7

−5.26

(− 0.9%)

−18.63 to 10.68

(− 3.1 to 1.8%)

PRP

PC100 platelet counter vs. Sysmex XP-300

250

251.6

1.56

(0.6%)

−10.32 to 12.32

(− 4.2 to 4.9%)

3600

3432.5

− 67.46

(−1.9%)

− 99.93 to − 30.17

(− 2.9% to − 0.9%)