Skip to main content
Fig. 3 | Thrombosis Journal

Fig. 3

From: Laboratory evaluation of a new integrative assay to phenotype plasma fibrinolytic system

Fig. 3

Summary of intra and inter-run precision for GFC. The line dividing the box into two parts represents the median of the data. The ends of the box show the upper (Q3) and lower (Q1) quartiles. The extreme line shows the highest and lowest value excluding potential outliers according to the following formula: Q3 + 1.5xIQR (maximum) and Q1-1.5xIQR (minimum). Black dots beyond the extreme line show potential outliers, grey dots represent individual data. Mean (m), standard deviation (SD) and relative standard deviation (RSV) are given as supplementary indicators

Back to article page