Successful retrieval of dislocated inferior vena cava filter using double vascular sheaths docking technology: case report

Abstract

Background

Dislocation of inferior vena cava filter (IVCF) is a rare complication with potential IVC perforation and other life-threatening risks requiring early diagnosis and in-time retrieval. Most of dislocation IVCF in the past have been shelved or removed by open surgery. It is very difficult to retrieve the filters by interventional technology.

Case presentation

Here we report a 49-year-old man suffering from dislocation of IVCF implanted due to deep vein thrombosis (DVT) in the right femoral vein. Successful retrieval of the IVCF using double sheaths docking technique was done soon after confirmation of the dislocation. Importance of monitoring and early detection of dislocation of IVCF should be emphasized to avoid further complications.

Conclusions

The double vascular sheaths docking technique can be considered as a preferential option in difficult operative situation.

Background

DVT in lower extremities occurs under conditions of venous endothelial damage, blood hypercoagulation, and stasis. The main risks of DVT include early fatal pulmonary embolism (PE) and late pulmonary hypertension. For patients with anticoagulation contraindications and recurrent DVT, IVCF should be implanted to prevent fatal PE. Furthermore, monitoring on IVCF status is critical to guarantee its’ thrombus capture effect and reduce filter-derived complications.

IVCF-derived adverse events have been reported as placement issues (45.1%), IVC penetration (29.9%) and IVC filter fracture (27.1%) [1,2,3,4,5]. At the same time, related data suggested a high rate of serious complications after temporary IVCF implantation, calling for the urgency of strictly following indications of IVCF and carefully monitoring. Early diagnosis and in-time retrieval are important to avoid severe complications.

We report an insidious case of IVCF leg dislocation, potentially leading to perforation of the vena cava. Double sheaths docking technology was used to successfully retrieved the IVCF, verifying the importance of timely handling of IVCF migration.

Case presentation

A 49-year-old man with DVT in right femoral vein following ruptured cerebral aneurysm underwent retrievable IVCF (Denali, Bard, USA) implantation to prevent fatal PE. Pharmoco-mechanical thrombectomy was performed with angiojet system (Boston Scientific, MA, USA). During the surgery, a leg of the filter was found to be dislocated (Fig. 1A), which may cause perforation of the vena cava and retroperitoneal hematoma [6, 7]. With the help of disposable biopsy forceps (KLF-A, Hangzhou, China) (Fig. 1B), the dislocated filter leg was taken into an 8F vascular sheath (Fig. 1C). Then routine filter retrieval (13F) procedures were carried out via jugular vein approach but failed to trap the filter hook due to filter tilt and hook apposition to the cava wall. The “Loop-Snare Technique” [8] was used to trap the filter hook with continuous push of the 13F sheath downwards to the 8F one to complete docking of the two vascular sheaths, resulting in successful retrieval of the filter (Fig. 1D). During the retrieval procedures, the entire filter was kept in the lumen of the two vascular sheaths to avoid possible injury to the vena cava (Fig. 1E-F).

Fig. 1
figure 1

Combination of double sheaths docking technique and Loop-Snare technique to retrieve the IVC filter. (A) One leg of the filter is dislocated and bent cephalad. (B-C) use of disposable biopsy forceps to retrieve the dislocated filter leg into an 8F vascular sheath, (D) push the 13F vascular sheath downwards to complete the docking of the two vascular sheaths and retrieve the filter, (E-F) model for double sheaths docking technique to retrieve the filter. * 8F sheath; # 13F sheath; → 0.035 in. hydrophilic guide wire

Discussion

IVCF is an absolutely therapeutic choice for avoiding fatal PE when conventional anticoagulation is contraindicated or deemed ineffective [9]. A retrievable filter was our choice for the patient on considering the patient’s age and relatively good clinical expectation. In view of functional design, the axial stability of the temporary filter in the inferior vena cava is worse than the permanent one [2, 3], so postoperative monitoring of the filter is the key to maintain a good thrombus interception effect and avoid filter-related complications. Related studies on the long-term complications of filter concluded that most adverse events related to IVCF were device migration (35.6%), fractured parts embolization (15.9%) and IVC perforation (7.6%) [2, 3]. In our vascular center, approximately 250 patients receive IVCF implantation due to suffering DVT of lower extremities each year. The most common adverse events were the poor axial tilt and the displacement of the IVCF, resulting in the difficulty in IVCF retrieval. Hemodynamics change of the IVC due to ventilation and cardiopulmonary resuscitation has been considered as the main factors contributing to filter dislocation [10, 11]. Previous studies have reported multiple cases of cardiac tamponade caused by filter dislocation [9, 12,13,14,15,16] with serious consequences. Therefore, complications related to filter dislocation require prompt treatment.

This case revealed filter leg dislocation and axial migration of the filter. The dislocated filter leg may not only causes perforation of the inferior vena cava, resulting in retroperitoneal hematoma [17], and causes injure to the intestine, pancreas, aorta [18], but also leads to leg fracture of filter during retrieving of the filter. Nishikawa T reported that the dropped part of fractured filter leg was difficult to retrieve and had to be left in the vena cava [19]. However, recent studies demonstrated the safety and technically feasibility of endovascular retrieval of the filter that penetrated adjacent intestine [18] and adjacent vertebrae [20]. In our case, we successfully retrieve the dislocated IVCF combined with the filter tilting exceeds 25 degrees, which is often regarded as one of the important reasons for the difficulty of retrieval [21], the possible reasons are as follows: (1) The close follow-up allowed us to discover the displacement of the filter promptly, and the foot of the filter has not penetrated the vena cava; (2) Double vascular sheaths docking technology is beneficial to correct the axial direction of the filter. To our knowledge, this is the first case reporting the double sheaths docking to retrieve the dislocated filter, the docking with double vascular sheaths to retrieve the filter has the following advantages: (1) the dislocated filter leg was kept with the head-side orientation, avoiding damage to the vena cava by the barbed leg (2) the dislocated filter leg is protected by the vascular sheath, limiting the movement range of the leg, and avoiding the possibility of secondary dislocation of the leg. Recent study reviewed the complications of IVCF malfunction including the penetration, fracture, or migration of the device, and assessed the result of open surgery retrieval and endovascular retrieval. Comparing to the open surgery retrieval, endovascular retrieval of IVCF has a significantly lower likelihood of thromboembolic complication, and a trend to lower infectious complications, mortality, hospital costs, however, with no statistical significance [22]. Therefore, endovascular retrieval of dislocated IVCF may have obvious advantages compared with open surgery due to its minimal invasiveness and safety. This report aims to highlight the importance of postoperative filter monitoring to avoid serious filter-derived complications. The double vascular sheaths docking technique can be considered as a preferential method to retrieve dislocated IVCF.

Conclusion

Dislocation of the IVCF is a rare and serious complication, calling for our sufficient attention. Monitoring on IVCF status is critical to guarantee its’ thrombus capture effect and reduce filter-derived complications. The double vascular sheaths docking technique can be considered as a superior method.

Availability of data and materials

All data analyzed during our study are included within the published article.

Abbreviations

IVCF:

Inferior vena cava filter

DVT:

Deep vein thrombosis

PE:

Pulmonary embolism

References

  1. Stavropoulos SW, Sing RF, Elmasri F, Silver MJ, Powell A, Lynch FC, et al. The DENALI Trial: an interim analysis of a prospective, multicenter study of the Denali retrievable inferior vena cava filter. J Vasc Interv Radiol. 2014;25:1497–1505, 1505 e1491.

    Article  Google Scholar 

  2. Ayad MT, Gillespie DL. Long-term complications of inferior vena cava filters. J Vasc Surg Venous Lymphat Disord. 2019;7:139–44.

    Article  Google Scholar 

  3. Wang SL, Siddiqui A, Rosenthal E. Long-term complications of inferior vena cava filters. J Vasc Surg Venous Lymphat Disord. 2017;5:33–41.

    Article  Google Scholar 

  4. Zhou D, Spain J, Moon E, McLennan G, Sands MJ, Wang W. Retrospective review of 120 celect inferior vena cava filter retrievals: experience at a single institution. J Vasc Interv Radiol. 2012;23:1557–63.

    Article  Google Scholar 

  5. Stoneham GW, Burbridge BE, Millward SF. Temporary inferior vena cava filters: in vitro comparison with permanent IVC filters. J Vasc Interv Radiol. 1995;6:731–6.

    Article  CAS  Google Scholar 

  6. Miller CL, Wechsler RJ. CT evaluation of Kimray-Greenfield filter complications. AJR Am J Roentgenol. 1986;147:45–50.

    Article  CAS  Google Scholar 

  7. Greenfield LJ. Perforation of the inferior vena cava with aortic and vertebral penetration by a suprarenal Greenfield filter. Radiology. 1990;175:287.

    Article  CAS  Google Scholar 

  8. Rubenstein L, Chun AK, Chew M, Binkert CA. Loop-snare technique for difficult inferior vena cava filter retrievals. J Vasc Interv Radiol. 2007;18:1315–8.

    Article  Google Scholar 

  9. Mocavero PE, Stagliano A, Coppola M, Langella S, Conte A, Pappalardo F, et al. An unusual cause of cardiogenic shock: tricuspid regurgitation and right ventricular perforation due to vena cava filter migration. HSR Proc Intensive Care Cardiovasc Anesth. 2012;4:47–50.

    CAS  PubMed  PubMed Central  Google Scholar 

  10. Laborda A, Kuo WT, Ioakeim I, De Blas I, Malve M, Lahuerta C, et al. Respiratory-induced Haemodynamic changes: a contributing factor to IVC filter penetration. Cardiovasc Intervent Radiol. 2015;38:1192–7.

    Article  Google Scholar 

  11. Chalhoub V, Richa F, Hachem K, Slaba S, Yazbeck P. Contributing factors to inferior vena cava filter migration. Cardiovasc Intervent Radiol. 2015;38:1676–7.

    Article  Google Scholar 

  12. Lorch H, Welger D, Wagner V, Hillner B, Strecker EP, Herrmann H, et al. Current practice of temporary vena cava filter insertion: a multicenter registry. J Vasc Interv Radiol. 2000;11:83–8.

    Article  CAS  Google Scholar 

  13. Millward SS. Buckling of the tethering catheter causes migration of a temporary caval filter to the right atrium. Clin Radiol. 2000;55:652–3.

    Article  CAS  Google Scholar 

  14. Rossi P, Arata FM, Bonaiuti P, Pedicini V. Fatal outcome in atrial migration of the Tempofilter. Cardiovasc Intervent Radiol. 1999;22:227–31.

    Article  CAS  Google Scholar 

  15. Bovyn G, Gory P, Reynaud P, Ricco JB. The Tempofilter: a multicenter study of a new temporary caval filter implantable for up to six weeks. Ann Vasc Surg. 1997;11:520–8.

    Article  CAS  Google Scholar 

  16. Sakai Y, Masuda H, Arai G, Kobayashi T, Kageyama Y, Kihara K. Accidental dislocation of an intracaval temporary filter into the heart in a case of renal cell carcinoma extending into the vena cava. Int J Urol. 2006;13:1118–20.

    Article  Google Scholar 

  17. Seita J, Sakakibara Y, Jikuya T, Shigeta O, Nakata H, Tsunoda H, et al. Surgical management of a penetrated greenfield inferior vena cava filter. Thorac Cardiovasc Surg. 2001;49:243–4.

    Article  CAS  Google Scholar 

  18. Joe WB, Larson ML, Madassery S, Devane AM, Holly BP, Sing RF, et al. Bowel penetration by inferior vena cava filters: feasibility and safety of percutaneous retrieval. AJR Am J Roentgenol. 2019;213:1152–6.

    Article  Google Scholar 

  19. Nishikawa T, Oka T, Shioyama W, Kashimoto K, Yasui T, Oboshi M, et al. Fracture of a retrievable inferior vena cava filter placed for Cancer-associated thrombosis in a patient with malignant lymphoma. Intern Med. 2020;59:673–6.

    Article  Google Scholar 

  20. McCafferty BJ, Joe WB, Meh Chu G, Devane AM, Madassery S, Gunn AJ. Penetration of bone by inferior vena cava filters: safety and technical success of percutaneous retrieval. Curr Probl Diagn Radiol. 2021;50:662–4.

    Article  Google Scholar 

  21. Rosenthal D, Wellons ED, Hancock SM, Burkett AB. Retrievability of the Gunther tulip vena cava filter after dwell times longer than 180 days in patients with multiple trauma. J Endovasc Ther. 2007;14:406–10.

    Article  Google Scholar 

  22. Kim TI, Abougergi MS, Guzman RJ, Ochoa Chaar CI. Costs and complications of hospital admissions for inferior vena cava filter malfunction. J Vasc Surg Venous Lymphat Disord. 2021;9:315–20 e314.

    Article  Google Scholar 

Download references

Acknowledgements

None.

Funding

This study was supported by the National Key Research and Development Program of China (No. 2017YFC1104100).

Author information

Authors and Affiliations

Authors

Contributions

Mi Zhou and Yongquan Gu designed the methods, analyzed the data and results, Mi Zhou and Lixing Qi wrote the manuscript and prepared figures. All authors have read and agreed to the published version of the manuscript.

Corresponding author

Correspondence to Yongquan Gu.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the Ethics Committee of Xuanwu Hospital affiliated with Capital Medical University. Written patient consent for participation was obtained.

Consent for publication

Written patient informed consent for participation was obtained. All authors have read and agreed to the published version of the manuscript.

Competing interests

The authors declare that they have no conflict of interest.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhou, M., Qi, L. & Gu, Y. Successful retrieval of dislocated inferior vena cava filter using double vascular sheaths docking technology: case report. Thrombosis J 19, 56 (2021). https://doi.org/10.1186/s12959-021-00309-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12959-021-00309-3

Keywords