Comparison of different laboratory tests in the evaluation of hemorrhagic risk of patients using rivaroxaban in the critical care setting: diagnostic accuracy study

Abstract

Background

Rivaroxaban is a direct oral anticoagulant designed to dispense with the necessity of laboratory monitoring. However, monitoring rivaroxaban levels is necessary in certain clinical conditions, especially in the critical care setting.

Methods

This is a diagnostic accuracy study evaluating sensitivity and specificity of prothrombin time (PT), activated partial thromboplastin time (aPTT), and Dilute Russell viper venom time (dRVVT), to evaluate the hemorrhagic risk in patients taking rivaroxaban. The study used a convenience sample of 40 clinically stable patients using rivaroxaban to treat deep vein thrombosis or atrial fibrillation admitted in a private hospital in Brazil, compared to a group of 60 healthy controls. The samples from patients were collected two hours after the use of the medication (peak) and two hours before the next dose (trough).

Results

The correlation with the plasmatic concentration measured by anti-FXa assay was higher for PT and dRVVTS. The PT and aPTT tests presented higher specificity, while dRVVT was 100% sensible.

Conclusions

There was a strong correlation between the tests and the plasma concentration of the drug. Additionally, our results demonstrated the potential use of dRVVT as a screening test in the emergency room and the need of a second test to improve specificity.

Background

The traditional anticoagulant drugs that exist for the prevention and treatment of venous thromboembolism (VTE), prevention of cerebral vascular accident in patients with atrial fibrillation (AF), and secondary prevention in patients with acute coronary syndrome (ACS) require constant laboratory monitoring, which is sometimes burdensome and inconvenient for the patient. Rivaroxaban is a direct oral anticoagulant (DOAC) that dispenses with this type of control. It is an antithrombotic drug, that acts directly inhibiting activated factor X, impeding the generation of thrombin, and consequently preventing the formation of clots. It also has the advantage that it can be administered in a single daily dose [1].

Rivaroxaban has high bioavailability after oral administration, with a maximum peak of action at around 1.5 to 2 h after use (peak action), a mean half-life of between 5 and 9 h in young patients and 12 to 13 h in those aged over 75 years. It is eliminated in two ways: two thirds are metabolized by the liver (via CYP3A4 and CYP2J2) without any circulating active metabolite identified, and one third is excreted unaltered in the urine [2, 3].

Although direct oral anticoagulants have been designed to dispense the necessity of laboratory monitoring, the literature has demonstrated that this monitoring is potentially useful in certain conditions [2, 4, 5].

So far, it is known that rivaroxaban can prolong the times in the conventional clotting tests used to evaluate the risk of hemorrhage, such as prothrombin time (PT), activated partial thromboplastin time (aPTT), and thrombin time (TT), but it has not yet been possible to establish a therapeutic window of clinical interest that can be linearly correlated with the plasma concentration of the drug for these tests. The only two specific tests available to date, and that present correlation with the plasma concentration of the drug, are anti-FXa assay (indirect method) [1, 4], and liquid chromatography/mass spectrometry (direct method).

More recently, some authors have considered PT [6, 7] and dilute Russell viper venom time (dRVVT) [8,9,10,11] as the most promising tests for this type of monitoring. dRVVT evaluates only the common laboratory coagulation pathway test (factors X, V, II and I) after activation of factor X by Russell viper venom, minimizing the possibility of interference of dysfunction of the other clotting factors. There are also two different types of reagents in terms of the concentration of phospholipid used: the screen test (lowest concentration of phospholipid, dRVVTS) and the confirm test (highest concentration, dRVVTC) to evidence the presence of lupus anticoagulant. Exner et al., suggested that dRVVT could possibly be used to detect and maybe determine the plasmatic concentration of many anticoagulants, including rivaroxaban [11]. Douxfils et al. state also that the “russell viper venom time test” allows a rapid estimation of the intensity of anticoagulation mediated by rivaroxaban, although the authors did not differentiate the drugs type [12].

Altman and Gonzalez published a study in which they proposed that Russell’s viper venom is the most sensitive for identifying patients at risk of hemorrhage or exhibiting low anticoagulant effect, but they emphasize the need for other studies, to establish the sensitivity of other methods and identify cut-off values [8].

Objective

The primary objective of our work was to compare the sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of PT, aPTT and dRVVT, in order to exclude qualitatively the plasma concentrations that are relevant in critical care. The secondary objective was to correlate PT, aPTT, dRVVT (screen and confirm) assays with plasmatic concentration of the drug measured by the anti-Xa methodology in the same patients.

Methods

Study design, setting and ethics

This is a diagnostic observational study, conducted at a private hospital in São Paulo, Brazil, with a convenience sample of stable patients using the drug rivaroxaban.

The Institutional Review Board approved the protocol in advance under CAAE number: 34661614.1.0000.0071. All the subjects signed an informed consent form to perform blood tests and to participate in the study.

Subjects

The subjects of this study are all patients consecutively admitted in the period of September 22, 2014 to December 21, 2015, in daily use of rivaroxaban at daily doses of 10, 15 or 20 mg daily (single dose), due to diagnosed thrombosis of the lower limbs, or those at risk of embolism due to atrial fibrillation. Patients with creatinine clearance lower than 15 mL/min or using rivaroxaban twice-daily regiment were excluded.

Plasma of 60 healthy controls without any known defects in the blood coagulation was used to estimate the normal range for the PT, aPTT and dRVVT screen and confirm assays.

Laboratorial analytical methods

The patients’ blood samples were collected on two separate occasions: two hours after the use of the medication (“peak” moment) and two hours before the next dose (“trough” moment). The samples were collected according to the norms recognized by the Clinical Laboratory of the hospital, in test tubes containing 3.2% sodium citrate anticoagulant (Sarstedt, Newton, NC). Conventional PT tests were performed with the reagent STA Neoplastine CI Plus 10 and aPTT with STA Cephascreen 4 and processed on a Stago STA-R Evolution coagulation analyzer (Stago, Asnières-sur-Seine, France).

The Russell viper venom test was processed with the reagent STA Staclot DRVV (Screen and Confirm; Stago, Asnières-sur-Seine, France), at different phospholipid concentrations; one low, designated dRVVTS, and the other with a high concentration, designated dRVVTC.

The evaluation of anti-FXa assay was performed with the reagent STA-Liquid Anti-FXa with specific calibrator for rivaroxaban (Stago, Asnières-sur-Seine, France).

Definitions

In this study, the following definitions were used:

  • ≤ 30 ng/mL as the plasma concentration cut-off defined as safe for invasive procedures [13, 14];

  • ≤ 50 ng/mL as the plasma concentration indicating moderate risk cut-off defined by Lim et al. [15] and Levy et al. [16];

  • ≤ 100 ng/mL as the concentration that may lead to thrombolysis in ischemic stroke [14, 17];

  • sensitivity as the proportion of individuals with a higher plasma concentration than the threshold (30 ng/mL, 50 ng/mL or 100 ng/mL) and a test result above normality;

  • specificity as the proportion of individuals with a plasma concentration lower than the threshold (30 ng/mL, 50 ng/mL or 100 ng/mL) and a test result within normality range;

  • positive predictive value (PPV) as the proportion of individuals with a positive test result who actually present plasma concentrations higher than the threshold (30 ng/mL, 50 ng/mL or 100 ng/mL);

  • negative predictive value (NPV) as the proportion of individuals with a negative test result who do really present plasma concentrations lower than the threshold (30 ng/mL, 50 ng/mL or 100 ng/mL);

  • correlation grades: coefficients < 0.2 meaning very weak correlation; 0.2 to 0.39 as weak correlations; 0.40 to 0.59 as moderate; 0.60 to 0.79 strong correlation; > 0.80 very strong [18].

Statistical analysis

The plasmatic concentrations were described as means, standard deviation and 95% confidence intervals (CI). Spearman’s correlations were calculated between the anti-FXa assay values (above the lower limit of quantitation, LOQ, 25 ng/mL) and the laboratory tests of interest. The results were illustrated using scatter plots. The analyses were performed with the software GraphPad Prism 5 (GraphPad Software, San Diego, CA), considering a level of significance of 0.05. The results from healthy subjects were analyzed using the EP Evaluator 11.1.0.26 (Data Innnovations, LLC) software program. EP Evaluator follows the recommendations of the CLSI C28-A (Clinical and Laboratory Standards Institute guideline) [19]. The normal ranges were expressed in the 90% CI. The diagnostic test evaluation was calculated using the free online MEDCALC easy-to-use statistical software. CI for sensitivity and specificity are “exact” Clopper-Pearson CI; and, for the predictive values, the standard logit CI given by Mercaldo et al. [20] were used.

Results

Study group

In the study period, 40 patients required the use of anticoagulation and were included and evaluated (Table 1).

Table 1 Patients’ baseline data

Plasmatic concentration

The plasmatic concentration calculated by anti-FXa assay were as expected in trough, mainly less than the LOQ (25 ng/mL), for all dosages. In peak time, the results were: 63.4 ± 61.0 ng/mL (mean ± standard deviation, SD, with 95% CI: 24.7–102.1) for 10 mg; 142.5 ± 93.2 ng/mL (CI: 97.6–187.4) for 15 mg, and 203.4 ± 85.6 ng/mL (CI: 137.6–269.2) for 20 mg.

Correlation between different methods with plasmatic concentration

All results of PT, dRVVT screen and confirm (Figs. 1, 2 and 3) presented a strong positive correlation with anti-FXa assay (r ≥ 0.60). The correlation of aPTT (Fig. 4) was only moderate (r = 0.53).

Fig. 1
figure 1

Correlation between anti factor X (anti-FXa) assay and prothrombin time (PT). Dotted lines indicate the normal range, determined from the control group

Fig. 2
figure 2

Correlation between anti factor X (anti-FXa) and dilute Russell viper venom time (dRVVT) screen assays. Dotted lines indicate the normal range, determined from the control group

Fig. 3
figure 3

Relationship between anti factor X (anti-FXa) and dilute Russell viper venom time (dRVVT) confirm assays. Dotted lines indicate the normal range, determined from the control group

Fig. 4
figure 4

Correlation between anti factor X (anti-FXa) assay and activated partial thromboplastin time (aPTT). Dotted lines indicate the normal range, determined from the control group

Normal range

Table 2 shows the normal range calculated for TP, aPTT, dRVVs and dRVVc.

Table 2 Normal range calculated using samples from 60 health volunteers

Sensitivity, specificity, PPV and NPV

Rivaroxaban concentration < 30 ng/mL was observed in 38/80 samples; < 50 ng/mL, in 46/80; and <100 ng/mL, in 57/80. Figs. 5, 6, 7, 8 and Tables 3, 4, 5 show the results dispersion and the performance of different tests according to different thresholds. It is possible to observe that, as the threshold increases, the TP and aPTT sensitivity also increases and specificity is reduced. dRVVs and dRVVc had the highest sensitivity regardless of the threshold adopted.

Fig. 5
figure 5

Prothrombin time (PT; seconds) distribution according to rivaroxaban plasmatic concentration. Normal range is represented by the gray band with a lower limit of 12.6 s (90% confidence interval, CI: 12.37–12.85) and upper limit of 15.2 s (90% CI: 14.96–15.44)

Fig. 6
figure 6

Activated partial thromboplastin time (aPTT), (seconds) distribution according to rivaroxaban plasmatic concentration. Normal range is represented by the gray band with a lower limit of 26.8 s (90% confidence interval, CI: 26.1–27.5) and a upper limit of 35.6 s (90% CI: 34.7–36.6)

Fig. 7
figure 7

Dilute Russell viper venom time screen (dRVVTS) (seconds) distribution according to rivaroxaban plasmatic concentration. Normal range is represented by the gray band with a lower limit of 36.5 s (90% confidence interval, CI: 36.2–36.9) and upper limit of 40.2 s (90% CI: 39.9–40.6)

Fig. 8
figure 8

Dilute Russell viper venom time confirm (dRVVTC) (seconds) distribution according to rivaroxaban plasmatic concentration. Normal range is represented by the gray band with a lower limit of 35.3 s (90% confidence interval, CI: 34.8–35.8) and upper limit of 41.0 s (90% CI: 40.4–41.5)

Table 3 Performance of four different tests for rivaroxaban monitoring (threshold: 30 ng/mL)
Table 4 Performance of four different tests for rivaroxaban monitoring (threshold: 50 ng/mL)
Table 5 Performance of four different tests for rivaroxaban monitoring (threshold: 100 ng/mL)

Discussion

Among the direct oral anticoagulants (DOACs), rivaroxaban has been widely used in medical practice, particularly because it dispenses with the necessity of laboratory control in the majority of patients [4, 5, 8]. However, in conditions where a laboratory evaluation is desirable, it has been demonstrated by many authors [1] that the test of choice, due to its high specificity, is the plasmatic concentration obtained by the chromogenic anti-FXa assay with specific calibrator for rivaroxaban.

According to Baglin et al. [21], three methodologies can be additionally used for rivaroxaban monitoring: activated partial thromboplastin time (aPTT), prothrombin time (PT) and plasmatic concentration determination. Recent data pointed dRVVT as an additional tool [8, 9].

Levy et al. [10] cited that is very important to know when the last dose of the direct oral anticoagulant was taken by the patient in order to determine whether the levels are likely to increase or fall over time. There are some situations, however, in that obtaining this information is impossible, and this happens frequently in the emergency setting [14]. Besides, some patients might metabolize anticoagulant drugs differently. Therefore, it is important to know the performance of the evaluation techniques that are available in the laboratory, so that a correct approach can be used in the emergency room. Because of that, this study sought to evaluate the correlation and performance of different tests regarding the plasmatic concentration of rivaroxaban.

The plasmatic concentrations measured by anti-FXa assay were comparable to Mueck et al. [22]. Our results also show that PT was most closely correlated to plasmatic concentration measured by anti-FXa assay, followed by dRVVTs, dRVVTc and aPTT. The PT and aPTT correlation to plasmatic concentration was strongly discussed by Francart et al. [23], depending on reagent type. Douxfils et al. studied the response of two PT reagents and correlated PT data to the plasma concentration measured by HPLC-MS/MS. Correlation was found to be very strong (0.86) [9]. The differences between theirs and our results can be attributed to the characteristics of the patients evaluated, the reagents used and the methodology used in the plasmatic concentration measurement.

Gosselin et al. [10] correlated dRVVT to plasmatic concentration obtained by the direct method of liquid chromatography/mass spectrometry. They used the Siemens LA2 and Precision Biologics DRVVT reagents and presented a higher correlation coefficient (0.85 and 0.88, depending on the reagent). As for the PT and aPTT assays, the differences found can be attributed to the reagents, methods and even patients evaluated. In our study, if we isolate the results obtained with patients using 20 mg/day (n = 18), the correlation coefficient with anti-FXa assay would be 0.82.

Ebner et al. [14] recently described two thresholds of plasma concentration that are relevant in situations of emergency (30 and 100 ng/mL). Levy et al. [16] added the threshold of 50 ng/mL in cases of patients with severe bleeding. Lim et al., in 2016, described the same cut-off of 50 ng/mL for rivaroxaban, apixaban and dabigatran [15]. For this reason, the diagnostic test evaluation in our study was conducted using the three thresholds (< 30 ng/mL, < 50 ng/mL and <100 ng/mL).

Sensitivity and specificity of PT and aPTT were threshold-dependent. PT is indeed frequently described as an adequate test for rivaroxaban monitoring [4, 5, 24, 25]. Lim et al., in 2016, using the same reagent for PT assay, demonstrated a sensitivity and NPV above 90% for the cut-off of 50 ng/mL [15]. These results corroborate our findings. However, the observed differences in specificity and PPV can be attributed to the group studied. In addition to this influence, we believe that the differences in diagnostic accuracy related to aPTT may depend on the reagent used.

dRVVT screen and confirm tests presented maximum sensitivity and NPV, i.e., 100%, regardless of the threshold used. This would undoubtedly the safest test for exclusion of plasmatic concentration associated to hemorrhagic risk when the results are normal. The number of false positive results in relation to the total of exams for dRVVT screen and confirm varied from 32/80 to 51/80. These results are similar to the literature [7, 26].

Gosselin et al. [27], Exner et al. [11] described dRVVT as highly sensible for the presence of rivaroxaban. In fact, the authors described the application of dRVVT for many DOACs. Douxfils et al. [12] discussed also that dRVVT could be useful to assess pharmacodynamics of DOACs, with the advantage that it is a single test applied to different DOACs.

Gosselin et al. [27] describe that anticoagulants cause false results not only in coagulometric tests, as in the investigation of lupus anticoagulant, for example, but also in chromogenic tests. The interference of rivaroxaban on dRVVT was also described by other authors [26, 28, 29]. It is important to highlight that the sensitivity and specificity of a quantitative test are dependent on the cut-off value above or below which the test is positive [30].

Based on our results, we suggest that the patients with hemorrhagic risk (those with dRVVT screen and confirm tests above normality, that have sensitivity and low specificity) be submitted to a second test with higher diagnostic specificity (anti-Xa test calibrated with rivaroxaban or, in the absence of this, PT). This would allow the reduction of false positives and inadequate therapeutic interventions.

The potential limitations of this study must be considered, specially the reduced and highly homogeneous and clinically stable sample of patients, which might not be the case of other emergency settings. This must be considered in light of the fact that PPV and NPV are highly dependent on the disease prevalence.

Conclusions

In conclusion, our findings showed that there is a correlation between the tests studied and the plasma concentration of rivaroxaban, confirming the existing description in the literature. In addition, as shown by the sensitivity and specificity results, our study suggests the applicability of the tests evaluated in the screening of plasmatic concentration associated to hemorrhagic condition in patients using rivaroxaban. However, it is crucial that the laboratory informs the attending physician about the diagnostic limitations of this group of tests in the evaluation of patients using anticoagulants, using an appropriate and clear note in the lab results report. Additionally, our results show a cut-off dependent behavior of the tests that must be better investigated in future studies.

Abbreviations

ACS:

Acute coronary syndrome

AF:

Atrial fibrillation

anti-FXa:

Anti factor X

aPTT:

Activated partial thromboplastin time

C:

Confirmatory

dRVVT:

Dilute Russell viper venom time

PT:

Prothrombin time

S:

Screen

TT:

Thrombin time

VTE:

Venous thromboembolism

References

  1. Tripodi A. The laboratory and the direct oral anticoagulants. Blood. 2013;121(2):4032–5.

    Article  CAS  PubMed  Google Scholar 

  2. Leung LLK. Direct oral anticoagulants: Dosing and adverse effects. UpToDate. 2016. Available from: http://www.uptodate.com/contents/direct-oral-anticoagulants-dosing-and-adverse-effects?source=search_result&search=Direct+oral+anticoagulants&selectedTitle=1%7E131. Accessed in 2017 (May 15).

  3. Kubitza D, Becka M, Mueck W, Zuehlsdorf M. The effect of extreme age, and gender, on the pharmacology and tolerability of rivaroxaban – an oral, direct factor FXa inhibitor. Blood. 2006;108:905. Available from: http://www.bloodjournal.org/content/108/11/905?sso-checked=true. Accessed in 2017 (May 15).

  4. Tripodi A. The laboratory and the new oral anticoagulants. Clin Chem. 2013;59(2):353–62.

    Article  CAS  PubMed  Google Scholar 

  5. Baglin T, Hillarp A, Tripodi A, Elalamy I, Buller H, Ageno W. Measuring oral direct inhibitors (ODIs) of thrombin and factor Xa: a recommendation from the subcommittee on control of anticoagulation of the scientific and standardisation Committee of the International Society on thrombosis and haemostasis. J Thromb Haemost. 2013;11:756–760.

  6. Tripodi A. Results expression for tests used to measure the anticoagulant effect of new oral anticoagulants. Thromb J. 2013;11(1):9.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  7. Gosselin RC, Adcock DM. The laboratory’s 2015 perspective on direct oral anticoagulant testing. J Thromb Haemost. 2016;14(5):886–93.

    Article  CAS  PubMed  Google Scholar 

  8. Altman R, Gonzalez CD. Simple and rapid assay for effect of the new oral anticoagulant (NOAC) rivaroxaban: preliminary results support further tests with all NOACs. Thromb J. 2014;12(1):7.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Douxfils J, Mani H, Minet V, Devalet B, Chatelain B, Dogné JM, et al. Non-VKA oral anticoagulants: accurate measurement of plasma drug concentration. Biomed Res Int. 2015;2015:345138.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Gosselin RC, Adcock Funk DM, Taylor JM, Francart SJ, Hawes EM, Friedman KD, et al. Comparison of anti-Xa and dilute Russell viper venom time assays in quantifying drug levels in patients on therapeutic doses of rivaroxaban. Arch Pathol Lab Med. 2014;138(12):1680–40.

    Article  PubMed  Google Scholar 

  11. Exner T, Ellwood L, Rubie J, Barancewicz A. Testing for new oral anticoagulants with LA-resistant Russells viper venom reagents. An in vitro study. Thromb Haemost. 2013;109(4):762–5.

    Article  CAS  PubMed  Google Scholar 

  12. Douxfils J, Chatelain B, Hjemdahl P, Devalet B, Sennesael AL, Wallemacq P, et al. Does the Russell viper venom time test provide a rapid estimation of the intensity of oral anticoagulation? A cohort study. Thromb Res. 2015;135(5):852–60.

    Article  CAS  PubMed  Google Scholar 

  13. Pernod G, Albaladejo P, Godier A, Samama CM, Susen S, Gruel Y, et al. Management of major bleeding complications and emergency surgery in patients on long-term treatment with direct oral anticoagulants, thrombin or factor-Xa inhibitors: proposals of the working group on perioperative haemostasis (GIHP) – march 2013. Arch Cardiovas Dis. 2013;106(6–7):382–93.

    Article  Google Scholar 

  14. Ebner M, Birschmann I, Peter A, Spencer C, Härtig F, Kuhn J, et al. Point-of-care testing for emergency assessment of coagulation in patients treated with direct oral anticoagulants. Crit Care. 2017;21(1):32.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Li MS, Chapman K, Swanepoel P, Enjeti AK. Sensivity of routine coagulation assays to direct oral anticoagulants: patient samples versus commercial drug-specific calibrators. Pathology. 2016;48(7):712–9.

    Article  Google Scholar 

  16. Levy JH, Ageno W, Chan NC, Crowther M, Verhamme P, Weitz JI. Subcommittee on control of anticoagulation. When and how to use antidotes for the reversal of direct oral anticoagulants: guidance from the SSC of the ISTH. J Thromb Haemost. 2016;14(3):623–7.

    Article  CAS  PubMed  Google Scholar 

  17. Steiner T, Böhm M, Dichgans M, Diener HC, Ell C, Endres M, et al. Recommendations for the emergency management of complications associated with the new direct oral anticoagulants (DOACs), apixaban, dabigatran and rivaroxaban. Clin Res Cardiol. 2013;102(6):399–412.

    Article  CAS  PubMed  Google Scholar 

  18. Evans JD. Straighforward statistics for the behavorial sciences. Califórnia: Brooks/Cole Publishing Company; 1996.

    Google Scholar 

  19. CLSI. Defining, establishing and verifying reference intervals in the clinical laboratory; approved guideline, third edition. CLSI document C28-A3. Wayne, PA: Clinical and Laboratory Standards Institute; 2008.

    Google Scholar 

  20. Mercaldo ND, Lau KF, Zhou XH. Confidence intervals for predictive values with an emphasis to case-control studies. Stat Med. 2007;26(10):2170–83.

    Article  PubMed  Google Scholar 

  21. Baglin T, Keeling D, Kitchen S. British Committee for Standards in haematology. Effects on routine coagulation screens and assessment of anticoagulant intensity in patients taking oral dabigatran or rivaroxaban: guidance from the British Committee for Standards in haematology. Br J Haematol. 2012;159(4):427–9.

    Article  CAS  PubMed  Google Scholar 

  22. Mueck W, Schwers S, Stampfuss J. Rivaroxaban and other novel oral anticoagulants: pharmacokinetics in healthy subjects, specific patient populations and relevance of coagulation monitoring. Thromb J. 2013;11(1):10.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Francart SJ, Hawes EM, Deal AM, Adcock DM, Gosselin R, Jeanneret C, et al. Performance of coagulation tests in patients on therapeutic doses of rivaroxaban a cross-sectional pharmacodynamic study based on peak and trough plasma levels. Thromb Haemost. 2014;111(6):1133–40.

    Article  CAS  PubMed  Google Scholar 

  24. Lippi G, Favaloro EJ. Recent guidelines and recommendations for laboratory assessment of the direct oral anticoagulants (DOACs): is there consensus? Clin Chem Lab Med. 2015;53(2):185–97.

    Article  CAS  PubMed  Google Scholar 

  25. Samuelson BT, Cuker A, Siegal DM, Crowther M, Garcia DA. Laboratory assessment of the anticoagulant activity of direct oral anticoagulants: a systematic Review. Chest. 2017;151(1):127–38.

    Article  PubMed  Google Scholar 

  26. Martinuzzo ME, Barrera LH, D’adamo MA, Otaso JC, Gimenez MI, Oyhamburu J. Frequent false-positive results of lupus anticoagulant tests in plasmas of patients receiving the new oral anticoagulants and enoxaparin. Int J Lab Hematol. 2014;36(2):144–50.

    Article  CAS  PubMed  Google Scholar 

  27. Gosselin R, Grant RP, Adcock DM. Comparison of the effect of the anti-Xa direct oral anticoagulants apixaban, edoxaban, and rivaroxaban on coagulation assays. Int J Lab Hematol. 2016;38(5):505–13.

    Article  CAS  PubMed  Google Scholar 

  28. Wong WH, Yip CY, Sum CL, Tan CW, Lee LH, Yap ES, et al. A practical guide to ordering and interpreting coagulation tests for patients on direct oral anticoagulants in Singapure. Ann Acad Med Singap. 2016;45(3):98–105.

    PubMed  Google Scholar 

  29. van Os GM, de Laat B, Kamphuisen PW, Meijers JC, de Groot PG. Detection of lupus anticoagulant in the presence of rivaroxaban using Taipan snake venom time. J Thromb Haemost. 2011;9(8):1657–9.

    Article  PubMed  Google Scholar 

  30. Lalkhen AG, McCluskey A. Clinical tests: sensitivity and specificity. Contin Educ Anaesth Crit Care Pain. 2008;8(6):221-223. Available from: https://www.google.com.br/url?sa=t&rct=j&q=&esrc=s&source=web&cd=1&ved=0ahUKEwjVyYKvpfLTAhXEg5AKHQTiCRYQFggpMAA&url=http%3A%2F%2Fserene.me.uk%2Fhelpers%2Fsensitivity-specificity.pdf&usg=AFQjCNFKq268clQ1-Hi7t0SRB7ergjDvuw. Accessed in 2017 (May 15).

Download references

Acknowledgements

The authors thank Patricia Logullo for editorial assistance during manuscript preparation.

Funding

This study received no funding.

Availability of data and materials

Please contact author for data requests.

Author information

Authors and Affiliations

Authors

Contributions

MPC, VFA, JCCG, NH and CLPM designed and coordinated the study. MPC, PBMCD analyzed and interpreted data and drafted the manuscript. MPC, VFA and JCCG carried out the assays, interpreted data and drafted the manuscript. PBMCD performed the statistical analysis. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Marjorie Paris Colombini.

Ethics declarations

Ethics approval and consent to participate

The Institutional Review Board approved the study protocol (number: 34,661,614.1.0000.0071). All the subjects signed informed consent forms to undergo blood tests and to participate in the study.

Consent for publication

Not applicable (the study will not be published with individual data).

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Colombini, M.P., Derogis, P.B.M.C., de Aranda, V.F. et al. Comparison of different laboratory tests in the evaluation of hemorrhagic risk of patients using rivaroxaban in the critical care setting: diagnostic accuracy study. Thrombosis J 15, 21 (2017). https://doi.org/10.1186/s12959-017-0140-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12959-017-0140-6

Keywords